Simultaneous phase and amplitude aberration sensing with a liquid-crystal vector-Zernike phase mask

Simultaneous phase and amplitude aberration sensing with a liquid-crystal vector-Zernike phase mask

David S. Doelman Fedde Fagginger Auer Leiden Observatory, Leiden University, P.O. Box 9513, 2300 RA Leiden, The Netherlands Michael J. Escuti Department of Electrical and Computer Engineering, North Carolina State University, Raleigh, NC 27695, USA Frans Snik Leiden Observatory, Leiden University, P.O. Box 9513, 2300 RA Leiden, The Netherlands
Abstract

We present an enhanced version of the Zernike wavefront sensor, that simultaneously measures phase and amplitude aberrations. The “vector-Zernike” wavefront sensor consists of a patterned liquid-crystal mask, which imposes phase on the point spread function core through the achromatic geometric phase acting with opposite sign on opposite circular polarizations. After splitting circular polarization, the ensuing pupil intensity images are used to reconstruct the phase and the amplitude of the incoming wavefront. We demonstrate reconstruction of the complex wavefront with monochromatic lab measurements and show in simulation the high accuracy and sensitivity over a bandwidth up to .

\journal

ol \ociscodes: (010.1080) Adaptive optics; (010.7350) Wave-front sensing; (350.1370) Berry’s phase; (160.3710) Liquid crystals;

The introduction of phase-contrast microscopy by F. Zernike [1, 2, 3, 4] was a revolution for the field of biological sciences.While a standard microscope measures intensity variations due to absorption, a phase-contrast microscope measures phase differences from variations in index of refraction by coupling the phase differences into intensity variations with a phase mask inside the microscope. Otherwise unseen transparent structures could now be imaged in great detail. Another application of the phase-contrast method is the Zernike wavefront sensor (ZWFS). The ZWFS has a similar phase mask with a dot that offsets the phase of the core of the point spread function (PSF) by . The optimal size of the dot is 1.06 , where is the pupil diameter and the wavelength. The phase-shifted PSF core interferes with the rest of the PSF and phase aberrations are converted to electric field amplitude variations in the subsequent pupil plane [4, 5, 6]. Thus, a phase-only aberration can be measured directly with one intensity measurement. Extremely sensitive systems like direct-imaging instruments for exoplanet detection use the ZWFS to minimize instrumental aberrations. Non-common path correction with the ZWFS has been successfully demonstrated on-sky [7] and picometer precision has been achieved in the lab for the Wide Field Infrared Survey Telescope (WFIRST) [8].
One major advantage of the Zernike wavefront sensor is that it is the most photon-efficient WFS with a sensitivity factor for photon noise of unity [6, 9]. Also, the ZWFS constitutes a simple implementation for any optical system that only requires a focal-plane mask and the ability to image the pupil plane. However, the dynamic range is limited, and the mask applies (wavelength-dependent) scalar phase shifts. In addition, a reference pupil image is needed for static amplitude aberration correction. Therefore, time-variable amplitude aberrations are reconstructed as phase aberrations. Static and dynamic amplitude variations can come from contamination of optics, degradation in optical performance, Fresnel propagation effects, and scintillation. Several systems that measure both amplitude and phase have been proposed, however these systems are complicated [10, 11] or require diversity in time, like the phase-shifting ZWFS, where a variable-phase Zernike mask is used to reconstruct the complex wavefront with four consecutive measurements [12].

In this paper we present a liquid-crystal version of the ZWFS, the "vector-Zernike" wavefront sensor (vZWFS) that enables the simultaneous measurement of both phase and amplitude aberrations. This simple upgrade only requires the replacement of the Zernike mask in addition to splitting circular polarization. The property of optimal photon efficiency is maintained by the vZWFS and a liquid-crystal focal plane mask applies achromatic geometric phase, yielding an improved broadband performance [13]. The geometric phase applied by the mask, , is independent of wavelength and only depends on the fast-axis orientation, , of the mask and the handedness of the circular polarization state of the incoming light [14]

(1)
Figure 1: Layout of the vector Zernike wavefront sensor. The intensity distribution of the pupils depends on the incoming phase and sign of the polarization state.

Writing complex fast-axis orientation patterns is enabled by liquid-crystal direct-write technology [15] and achromatizing the half-wave retardance is enabled by stacking self-aligning liquid-crystal layers to form a multi-twist retarder [16]. The precise writing capability and broadband performance of this technology has been verified in the lab for other optical elements [17, 14]. A Zernike mask manufactured with liquid-crystal technology simultaneously applies phase to the right-handed circularly polarized fraction of the PSF core and to the left-handed circularly polarized fraction of the PSF core. We split the light with opposite handedness with a quarter-wave plate and a polarizing beam-splitter as shown in Fig. 1.
To describe the resulting intensity in the two pupils, we adopt the formalism from N’Diaye et al. [5]. The electric field at the entrance pupil, , is defined as

(2)

Here P is the normalized amplitude such that is the pupil function. The variables and are amplitude aberration and the phase aberration for each position in the pupil plane. Following the derivation of N’Diaye et al. [5] and assuming , the intensity of the two pupils ( and ) can be written as

(3)
(4)

where is the convolution of the Fourier transform of the binary Zernike mask and the pupil plane electric field. If the aberrations are small, can be approximated by , using the pupil function instead of the pupil plane electric field ().
Splitting opposite circular polarization states introduces the factor of a half and creates two pupils with different intensities depending on the sign of . This is used to measure both the amplitude and the phase aberrations in a similar fashion as Wallace et al. [12]. They use the sum and the difference of four intensity measurements to directly calculate the amplitude and phase aberration. Here the phase is measured in the same way, only the amplitude is measured as the square root of the pupil intensity. The sum and the difference of the pupil intensities are given by

(5)
(6)

Using these equations and the identity , we solve for the amplitude P and the phase . The exact reconstruction is given by

(7)
(8)

Solving these equations can become numerically unstable for real wavefront sensor applications. Similar to N’Diaye et al. [5], we assume that the vector-Zernike wavefront sensor operates in the low-aberration regime, where dominates the reconstruction. Therefore, and for the phase reconstruction P = 1. Then we calculate P using the approximated values for ,

(9)
(10)

Measuring both P and allows us to estimate the aberrated electric field and approximate . This process is iterated and converges in the linear regime to the right , and . Note that the reconstruction algorithms do not require many complex calculations such as matrix multiplications to recover the complex wavefront. Therefore the reconstruction is suitable for very fast wavefront reconstruction.
To compare the effect of amplitude aberrations in the reconstruction for both the ZWFS and the vector-ZWFS, we perform numerical simulations of both systems with HCIPy [18]. We start with an unobstructed pupil with both phase and amplitude aberrations as shown in the left column of Fig. 2. The input phase aberration has a = 0.22 rad rms and 1.17 rad peak-to-valley (PtV). The normalized amplitude aberration has a = 0.6% rms ( = 1.3%) and 3.7% PtV (I: 7.2% PtV), where I is the intensity. We reconstruct the phase for the ZWFS using Eq. 14 in N’Diaye et al. [5]. For the vector-ZWFS we use three different methods to calculate the phase and the amplitude. The first method (Linear), uses modal-based wavefront reconstruction to reconstruct 200 Zernike mode coefficients of the phase and the amplitude uses Eq. 10. The second method ( order) uses Eq. 9-10 to reconstruct the aberrations directly from the intensity measurements with 5 iterations. The third method (Exact) uses Eq. 7-8 using and 5 iterations.

Figure 2: Comparison in the simulated reconstruction of both phase and amplitude between the Zernike wavefront sensor (ZWFS) and the vector-ZWFS. The phase aberration is shown in the top row, followed by the residual phase, i.e. the difference between the reconstructed and the input phase. The two bottom rows contain the same with amplitude aberrations. The different columns show the results for different measurement and reconstruction methods.

We compare the standard deviation of the residual phase between the ZWFS and the vZWFS and find that the non-linear algorithms of the vZWFS outperform the ZWFS. For the second row in Fig. 2, from left to right starting from the input, rad rms. The phase reconstruction of the classical ZWFS in the second column of Fig. 2 shows that the residual phase aberration in the pupil is caused by the amplitude aberration that it can not directly correct for. This is different from all three vZWFS reconstruction methods, where the residual phase is dominated by the inability to reconstruct high-frequency aberrations with the first 200 Zernike modes (Linear), approximations used in the phase calculation ( order), and the dynamic range of the vZWFS (Exact). No significant influence of the amplitude aberration is seen in the residual phase. The amplitude aberration measurements of the vZWFS are dominated by phase estimation errors. The residual amplitude aberrations from left to right, starting from the input, are rms.
For practical implementations we look into the influence of imperfect optics shown in Fig. 1 and noise on the performance of the vZWFS. Measuring two pupils increases read noise with a factor of 2 and combining the two pupil measurements suffers from flat-fielding effects. In addition, the vZWFS setup requires the use of a liquid-crystal Zernike mask, a quarter-wave plate (QWP) and a polarizing beam splitter (PBS). The vector-Zernike mask has an efficiency (the amount of light that acquires the geometrical phase) that depends on the retardance of the liquid-crystal film [16]. When the phase mask has an offset of the half-wave retardance, , the incoming unpolarized light that does not acquire the phase is split by the PBS and adds a background proportional to the pupil intensity. Similar contaminations happen with retardance deviations of the QWP and the rotation of the PBS. The consequent left pupil intensity distributions (identical to the right pupil) from an imperfect HWP, QWP and PBS respectively are given by

(11)
(12)
(13)

To first order, the difference of the pupil intensities is most affected by the extra terms. All terms proportional to drop out and the difference now is proportional to a term that reduces the response. The reduced response can be corrected with a gain factor from calibration. The sum of the pupil intensities is only affected by the deviation from half-wave retardance of the vector-Zernike mask, all other terms sum to the total intensity . The chromatic retardance can be controlled with liquid-crystal technology such that [16] for a large bandwidth, i.e. the leakage is < of the intensity.

As a proof of principle, we demonstrate the vector-Zernike wavefront sensor with a monochromatic setup. We manufactured a liquid-crystal vector-Zernike mask with a central spot diameter of m, printed on a glass substrate (1 inch diameter, 1 mm thick, BK7) with an effective pixel size of 2 m. The retardance was tuned to be at 633 nm with a single liquid-crystal layer. Fig. 3 b) shows the fabricated mask that does not contain any significant defects. The edges have a smooth fast-axis transition region from 0 to 45 of 2-3 micron. This can be mitigated by modifying for calculating and , although the effect is negligible. We test the performance of the vector-Zernike using the setup shown in panel a) of Fig. 3.

Figure 3: a) Layout of the vector-Zernike wavefront sensor setup. We generate a clean beam with a laser (633nm) and a pinhole. A spatial light modulator is operated in phase-mostly configuration with two polarizers and is used to generate phase aberrations with a complicated pattern (i.e. the Leiden University logo). This configuration also generates some amplitude aberrations. The light is focused on the vector-Zernike mask with a spot diameter of 75 m, corresponding to . The liquid-crystal orientation is shown in panel b) and a parallel polarizer microscopic image is shown in panel c). The detector image with aberrated pupils is shown panel d), showing some pupil overlap because of the splitting angle of the Wollaston.

We use a LC2002 transmissive spatial light modulator (SLM) from HoloEye in phase-mostly configuration to generate phase aberrations. In this configuration the rotation of the liquid-crystals in combination with the polarizers also generates amplitude aberrations. The Wollaston prism is a WPQ10 prism from Thorlabs and the QWP is an achromat (AQWP05M-600) from Thorlabs. Panel d) in Fig. 3 shows a raw image after applying a binary logo of Leiden University on the SLM. For the characterization of phase and amplitude reconstruction we minimize the influence of the system by combining measurements with positive and negative phase applied by the SLM. We mean-combine them to remove the non-SLM aberrations and we bin the image per 2 pixels to increase the signal to noise. We also take images without any phase applied on the SLM and use them for calculating the pupil intensity necessary for normalizing the aberrated images without taking out the vector-Zernike mask. Note that taking the average pupil intensity does not normalize the measurements correctly. Assuming , we solve Eq. 5 with , such that is given by

(14)
Figure 4: a) Reconstructed phase and b) reconstructed amplitude aberration using the vector-Zernike wavefront sensor.

When , the normalized quantity and we calculate the normalization factor of the pupil images by estimating and averaging over the pupil. We calculate the phase and amplitude from the two aberrated normalized pupils with Eqs. 9 - 10. Both measurements contain Fresnel diffraction effects from the out-of-plane diaphragm used as field stop. Determining the accuracy of the reconstruction is not possible as we do not know the details of this SLM.
One additional advantage of the liquid-crystal vector-Zernike wavefront sensor is that the phase is applied achromatically cf. Eq. 1. The classical ZWFS has decreased performance with increasing spectral bandwidth because of the chromatic phase of the focal plane mask in combination with the PSF scaling with wavelength. Both contribute to a suboptimal coupling of phase aberrations in the system to intensity variations in the ZWFS pupil plane [5]. We study the effect of the achromatic phase and the different reconstruction on the broadband performance of the vZWFS in simulation, taking into account the spatial scaling of the PSF with wavelength. For this, we exclude amplitude aberrations. Fig. 5 shows the results for both wavefront sensors up to bandwidth, i.e. . We assume a scaling for the applied phase of the classical Zernike mask centered around and is calculated at as well. We investigate the influence of only the achromatic phase by using the classical ZWFS reconstruction on one of the vZWFS pupils, the "achromatic ZWFS". Note that all other chromatic effects, including leakage from polarization optics, are not taken into account in this simulation. Fig. 5 shows that the vZWFS with exact reconstruction outperforms the classical ZWFS for all bandwidths. Similarly, the classical reconstruction with the "achromatic ZWFS" improves the broadband performance, in agreement with Bloemhof [13]. The gain of using only achromatic phase is not as significant as the gain from the reconstruction using two pupils. The reason is that the classical ZWFS uses a second-order reconstruction that is sensitive to the chromatic , while the vZWFS allows for an exact solution. Iterating the estimate of with the updated phase and amplitude estimates removes the bias introduced by the chromaticity.

Figure 5: Residual wavefront aberrations as function of bandwidth for both the classical ZWFS and the vZWFS. The "achromatic ZWFS" uses the classical ZWFS reconstruction on one pupil of the vZWFS.

Overall we show in simulation that the vZWFS can handle an increased bandwidth, up to .

We conclude that the liquid-crystal vector Zernike wavefront sensor (vZWFS) is a simple yet powerful upgrade of the classical ZWFS. We show that both amplitude and phase aberrations can be measured simultaneously by replacing the Zernike mask with a liquid-crystal version and splitting circular polarizations in two pupils. We demonstrate both in simulations and with a monochromatic lab setup an improved reconstruction of the complex wavefront with the vZWFS. Like the ZWFS, the vZWFS is the most photon-efficient wavefront sensor [6] and we demonstrate with simulations the accurate reconstruction for bandwidths up to . For direct-imaging of exoplanets the vZWFS can be installed a parallel WFS beam path, alternated with a coronagraph [7], or built into a coronagraph [8].

Funding. The research of DSD and FS leading to these results has received funding from the European Research Council under ERC Starting Grant agreement 678194 (FALCONER)

References

  • [1] F. Zernike, Monthly Notices of the Royal Astronomical Society 94, 377 (1934).
  • [2] F. Zernike, Physica 9, 686 (1942).
  • [3] F. Zernike, Physica 9, 974 (1942).
  • [4] J. Goodman, Introduction to Fourier Optics, Electrical Engineering Series (McGraw-Hill, 1996).
  • [5] M. N’Diaye, K. Dohlen, T. Fusco, and B. Paul, Astronomy & Astrophysics 555, A94 (2013).
  • [6] O. Guyon, The Astrophysical Journal 629, 592 (2005).
  • [7] A. Vigan, M. N’Diaye, K. Dohlen, J. Milli, Z. Wahhaj, J.-F. Sauvage, J.-L. Beuzit, R. Pourcelot, D. Mouillet, and G. Zins, arXiv preprint arXiv:1806.06158 (2018).
  • [8] F. Shi, B.-J. Seo, E. Cady, B. Kern, R. Lam, D. Marx, K. Patterson, C. M. Prada, J. Shaw et al., Proc. SPIE 10698, 106982O (2018).
  • [9] C. Paterson, Journal of Physics: Conf. Series. 139, 012021 (2008).
  • [10] K. Yamamoto, T. Matsuo, H. Imada, and M. Kino, Applied optics 54, 7895 (2015).
  • [11] E. Cuche, P. Marquet, and C. Depeursinge, Applied optics 38, 6994 (1999).
  • [12] J. K. Wallace, S. Rao, R. M. Jensen-Clem, and G. Serabyn, Proc. SPIE 8126, 81260F (2011).
  • [13] E. Bloemhof, Proc. SPIE 8999, 89991k (2014).
  • [14] M. J. Escuti, J. Kim, and M. W. Kudenov, Optics and Photonics News 27, 22 (2016).
  • [15] M. N. Miskiewicz and M. J. Escuti, Optics Express 22, 12691 (2014).
  • [16] R. K. Komanduri, K. F. Lawler, and M. J. Escuti, Optics Express 21, 404 (2013).
  • [17] C. Oh and M. J. Escuti, Optics Letters 34, 3637 (2009).
  • [18] E. H. Por, S. Y. Haffert, V. M. Radhakrishnan, D. S. Doelman, M. Van Kooten, and S. P. Bos, Proc. SPIE 10703, 1070342 (2018).
\bibliographyfullrefs

vector_zernikeWFS.bib

Comments 0
Request Comment
You are adding the first comment!
How to quickly get a good reply:
  • Give credit where it’s due by listing out the positive aspects of a paper before getting into which changes should be made.
  • Be specific in your critique, and provide supporting evidence with appropriate references to substantiate general statements.
  • Your comment should inspire ideas to flow and help the author improves the paper.

The better we are at sharing our knowledge with each other, the faster we move forward.
""
The feedback must be of minimum 40 characters and the title a minimum of 5 characters
   
Add comment
Cancel
Loading ...
321589
This is a comment super asjknd jkasnjk adsnkj
Upvote
Downvote
""
The feedback must be of minumum 40 characters
The feedback must be of minumum 40 characters
Submit
Cancel

You are asking your first question!
How to quickly get a good answer:
  • Keep your question short and to the point
  • Check for grammar or spelling errors.
  • Phrase it like a question
Test
Test description