Neutron spectroscopy with the Spherical Proportional Counter based on nitrogen gas

Neutron spectroscopy with the Spherical Proportional Counter based on nitrogen gas

E. Bougamont A. Dastgheibi J. Derre J. Galan G. Gerbier I. Giomataris M. Gros I. Katsioulas ioannis.katsioulas@cea.fr D. Jourde P. Magnier X.F. Navick T. Papaevangelou I. Savvidis G. Tsiledakis IRFU, Centre d’études de Saclay, Gif sur Yvette, France Queen’s University, Kingston, Canada Aristotle University of Thessaloniki, Thessaloniki, Greece LSM, CNRS/IN2P3, Université Grenoble-Alpes, Modane, France
Abstract

A novel large volume spherical proportional counter, recently developed, is used for neutron measurements. The pure gas is studied for thermal and fast neutron detection, providing a new way for neutron spectroscopy. The neutrons are detected via the and reactions. The detector is tested for thermal and fast neutrons detection with and neutron sources. The atmospheric neutrons are successfully measured from thermal up to several MeV, well separated from the cosmic ray background. A comparison of the spherical proportional counter with the current available neutron counters is also presented.

\pdfstringdefDisableCommands

1 Introduction

The Spherical Proportional Counter (SPC) (figure 1) is a novel concept with very promising features, among which is the possibility of easily instrumenting large target masses with good energy resolution and low energy threshold. The natural radial focusing of the spherical geometry allows collecting and amplifying the deposited charges by a simple and robust detector using a single electronic channel to read out large gaseous volumes (giomataris1, ; giomataris2, ; aune, ). The sensitivity of the detector is obtained using the Rn daughters decay to alpha particles in the energy range of . The low energy calibration of the detector from up to several tenths of keV is achieved by using X-rays and a UV lamp (giomataris3, ; andriamonje, ). The detector has been successfully used as a neutron detector based on the reaction (savvidis, ). At the present work the SPC has been optimized for operation with pure nitrogen (), which presents an interesting neutron converter, permitting neutron spectroscopy up to . The thermal and fast neutrons are detected via the reaction and fast neutrons with energy greater than 1.7 MeV via the reaction. The low background of the detector and the possibility to separate the ray pulses from proton and alpha particle pulses (through pulse shape analysis) increases the sensitivity for neutron detection. This detector can successfully measure very low thermal neutron fluxes, enabling measurements of neutron energy spectra up to several MeV at ground and underground level (savvidis, ).

Figure 1: Pictures of the SPC and the electronics setup in (a) the CEA, Saclay and (b) the Aristotle University of Thessaloniki.

2 The detector

The detector consists of a large spherical copper vessel, in diameter and thick. It is operated in a sealed mode, after being well pumped (up to ) and filled with a gas mixture at pressure of several hundred up to 5 bar. An outgassing in the order of is necessary for the amplification stability, because the presence of the in the drift volume induces electron attachment. A small stainless ball (sensor), usually of in diameter, fixed in the center of the spherical vessel by a stainless steel rod, acts as an electrode with positive high voltage (anode) in respect to the grounded vessel (cathode) and as a proportional amplification counter. Details about the design and principle of the SPC can be found at (giomataris3, ) where a complete description of the detector exists. In order to fulfill the requirements for operation with pure the ball had to be replaced by a smaller or ball. In addition, the metallic ball had been replaced by a silicon one to reduce the negative effects of occasional sparks. These improvements were necessary to achieve proportional amplification in pure , a process occurring at extremely high electric fields. Figure 2 shows a typical spectrum produced by a source at pressure, the obtained resolution () is reasonably good at about .

Figure 2: Spectrum of the source obtained with the spherical detector filled with nitrogen () gas at .

3 The neutron detection

The detector is filled with pure at pressure up to . Thermal neutrons interact with the nucleus via the reaction

The energy of the exothermic reaction is shared as kinetic energy between the and the proton with and . Fast neutrons can be detected via the

reaction up to . The cross section for the thermal neutron detection is gledenov () and for fast neutron detection is presented in figure 3. The endothermic reaction is the one with the most significant contribution to fast neutron detection for incident neutron energies greater than , since the cross section of the reaction is higher than that of the reaction after this energy value.

Figure 3: The cross sections of the reactions and for neutrons of energy up to (endfvII, ).

The neutron detection results are presented in figure 4 as the rise time versus amplitude of the pulse and the amplitude spectrum in ADC units (figure 4). The rise time of the pulse provides the depth of the ionization electrons produced in the gas and the amplitude the deposited energy. The gas used was pure at pressure. The detector was irradiated by a source, cased inside a polyethylene (PE) half box (that covered the back solid angle), placed away from the surface of the spherical detector. This technique was used to thermalize a part of the fast neutron spectrum of the source. The peak at belongs to thermal neutrons detected through the reaction and is well separated from the cosmic rays and the nuclear recoils. The fast neutrons are detected through both the and the reactions and appear to the right of the thermal peak. The recoil nuclei, which are produced by the fast neutron elastic scattering, appear in the region left (below ) from the thermal peak at lower energies.

Figure 4: Thermal neutron measurements after irradiation by a source. The gas used was of pure , the sensor was in diameter, the high voltage was set at providing a gain of . The measurements presented are (a) the rise time as a function of the amplitude, (b) the amplitude spectrum in ADC units. The detection energy threshold was .

The atmospheric neutrons have been also successfully measured with pure at . The thermal neutron peak and the fast neutrons measured are presented in figure 5 after 16 hours of acquisition.

Figure 5: Atmospheric neutron detection with pure gas, with a sensor of in diameter and a high voltage of . The measurements presented are (a) the rise time as a function of the amplitude (b) the amplitude spectrum in ADC units. The detection energy threshold was 75 keV and the rate in the region was . This rate is considered to be the total background rate (considering muon and neutrons incduced events) for fast neutron measurements.

4 Fast neutrons measured with and neutron sources and the wall effect

Long irradiation time measurements with () and () neutron sources have been performed, to study fast neutron detection. In figure 6 we present the measured energy distribution of the events produced by the fast neutrons of the source (figure 6. For the measurement, the detector was filled with pure at and a sensor with a silicon ball, in diameter, was used. The neutrons have a distribution with a maximum close to (figure 6). In the energy range of the neutrons, the reaction plays a dominant role (figure 3). In the energy spectrum (figure 6), we see the characteristic cross section peaks and the accumulation of signals at lower energies because of the wall effect (wall1, )(wall2, ). In order to measure the neutron energy, the produced particles from both and reactions must deposit all their energy in the drift volume. If the reaction takes place close to the vessel wall, it is possible for one of the charged particles or both of them to hit the wall and to lose a part of their energy. This is known as the wall effect and leads to wrong estimation of the neutron energy.

Figure 6: (a) Measured energy spectrum after irradiation by fast neutrons from a neutron source ( sensor ball, gas filling at ). The cross section coincidence peaks appear in the measured energy deposition spectrum. (b) source neutron energy spectrum (iso, ). Taking into account the activity of the source and its position relative to the SPC, the efficiency was estimated to be .

In figure 7 we present the measurements after irradiation by an neutron source. The sensor used was an metallic ball and pure gas at pressure. Since the neutrons of an source (figure 7) have higher energy compared to the neutrons of a source, most events are coming from the reaction. As in the previous figure, we can see the accumulation of signals at lower energies because of the wall effect and the neutron moderation by the surrounding material.

Figure 7: (a) Measured energy spectrum after irradiation with fast neutrons from a source ( sensor ball, gas filling at ). There is a translation of the spectrum to lower energies because of the wall effect. (b) source neutron energy spectrum (iso, ).

In case the detector being used as a neutron spectrometer and not as a simple neutron counter, the wall effect plays an important role. The wall effect depends on the range of the produced particles and on the size of the detector. In figure 8 we present the wall effect for the geometry of the SPC as a function of the sum of the ranges of the produced charged particles. The calculation was done using a simplified Monte Carlo simulation; we considered a homogeneous distribution of the interaction point inside the spherical volume and an isotropic, back-to-back emission of the reaction products. If any of the two fragments touches the detector shell, the event is accounted as ”lost due to wall effect”. The ranges depend on the neutron energy and on the gas pressure. As an example, for pure gas at and thermal neutrons, the range of the emitted proton is and the range of the is , resulting in of the events lost due to wall effect. The wall effect is more important for the fast neutrons since the range of protons is in pure at and the loss of events is about . For neutron energies higher than the reaction is the dominant one for the fast neutron detection. The ranges of the and the particles in for neutrons are approximately and respectively and the loss of events is .

Figure 8: Calculated fraction of recorded events (), with partial energy deposition due to the wall effect as a function of the range of the reaction products, for the SPC with a diameter of 130 cm.

5 Comparison with the commercial neutron counters.

There are many commercially available neutron counters (knoll, ) mainly in cylindrical shape, with different diameter and length. The tubes are filled with gas (he3, ) or gas (bf3, ) and the neutrons are detected via the and reactions. Both counter types, are very sensitive to thermal neutrons because of the positive Q-values and the high cross sections.

: and

: and

In comparison with the and , the has lower cross section for thermal neutrons, but this disadvantage can be covered sufficiently by the large amount of nitrogen atoms in the spherical counter. Thanks to the large volume of the sphere the wall effect is very small compared with that of the cylindrical tubes. Additionally the cost of the gas is very high and the is toxic. For fast neutrons up to the three isotopes, , and , have comparable cross sections (figure 9), but the fast neutron events in the cylindrical detectors are masked by the thermal ones. Their detection needs the help of moderators. Whereas the large volume of the spherical detector () with the large amount of gas (up to ) inside, provides the possibility of direct fast neutron spectroscopy. The use of instead of in the SPC is also suitable for fast neutron spectroscopy; compared to the gas it produces nuclear recoils in the same energy region as the thermal neutron peak, increasing the background.

6 Future developments

The first measurements revealed the capabilities of the spherical detector based on as a neutron counter and its potential as a simple neutron spectrometer. The performance of the detector is expected to be improved in the coming tests by considering the following modifications:

  • operation in higher gas pressure, thus increasing the detection efficiency and minimizing at the same time the wall effect

  • reduce the diameter of the electrode ball

  • use mixtures of with heavy gases like Ar and Xe. The addition of those gases will reduce the range of the reaction products, also limiting the wall effect

Figure 9: Comparison of the cross sections of the reactions , , and for fast neutrons of energy up to (endfvII, ).

Ongoing simulations show that particle discrimination through pulse shape analysis technique is possible. The separation is based on the fact that the range of protons in the gas is longer than the one of alpha particles of same energy. The radial distance between the start and end point of the particle trajectory (together with the longitudinal diffusion) reflects on the time characteristics of the corresponding pulses. The correlation of the observed rise time and pulse width with the pulse amplitude will allow the discrimination of from reactions and the rejection of reactions, thus simplifying the reconstruction of the neutron spectrum. Furthermore, the wall effect is much less important for the reaction, so it is expected to extend the range of the spectroscopy to higher neutron energies.

In order to exploit the spectroscopy capabilities of the spherical detector we are planning a series of measurements with monoenergetic neutron beams in facilities like CEA Cadarache or N.C.S.R. Athens.

Figure 10: Radially projected particle range as a function of the deposited energy, from GEANT4 (agostinelli, ) simulations for gas. The radial range reflects on the pulse characteristics: larger length will give larger rise time and pulse width. The events corresponding to alpha particles, protons or electrons are distributed in three distinguishable populations.

7 Conclusions

The SPC can be successfully used as a neutron detector, with pure gas, exploiting the and reactions. We were able to measure the thermal neutron flux from both neutron sources and atmospheric neutrons with success. As for the fast neutrons we were able to measure the and the neutrons and to show the fast neutron spectroscopy potential of the spherical detector, for neutron energies up to . The simple, robust design, the large volume and the possibility of operation at high pressure are important advantages that make the SPC an attractive alternative to the typical cylindrical counters.

Acknowledgments

This research/publication has been co-financed by the European Union (European Social Fund – ESF) and Greek national funds through the Operational Program ”Education and Lifelong Learning” of the National Strategic Reference Framework (NSRF) - Research Funding Program: “THALIS – HELLENIC OPEN UNIVERSITY- Development and Applications of Novel Instrumentation and Experimental Methods in Astroparticle Physics”.

References

  • (1) I. Giomataris, J. Vergados, Nuclear Instruments and Methods in Physics Research Section A 530 (3) (2004) 330 – 358. doi:http://dx.doi.org/10.1016/j.nima.2004.04.223.
  • (2) I. Giomataris et al, Nuclear Physics B - Proceedings Supplements 309 (1) (2006) 012031. doi:http://dx.doi.org/10.1016/j.nuclphysbps.2005.01.245.
  • (3) S. Aune et al, AIP Conf.Proc. 110–8 (2005) 785.
  • (4) I Giomataris et al, Journal of Instrumentation 3 (09) (2008) P09007. doi:doi:10.1088/1748-0221/3/09/P09007.
  • (5) S. Andriamonje et al, J. Phys. Conf. Ser. 179 (2009) 012003. doi:10.1088/1742-6596/179/1/012003.
  • (6) I. Savvidis et al, TAUP Conference 785 (2009) 110–8.
  • (7) Y. M. Gledenov, V. I. Salatski, P. V. Sedyshev, The 14n(n,p)14c reaction cross section for thermal neutrons, Zeitschrift für Physik A Hadrons and Nuclei 346 (4) (1993) 307–308. doi:10.1007/BF01292522.
  • (8) M.B. Chadwick et al, ENDF/B-VII.1 nuclear data for science and technology: Cross sections, covariances, fission product yields and decay data, Nuclear Data Sheets 112 (12) (2011) 2887 – 2996, special Issue on ENDF/B-VII.1 Library. doi:http://dx.doi.org/10.1016/j.nds.2011.11.002.
  • (9) R. Cervellati, A. Kazimierski, Wall effect in bf3 counters, Nuclear Instruments and Methods 60 (2) (1968) 173 – 178. doi:http://dx.doi.org/10.1016/0029-554X(68)90407-2.
  • (10) S. Shalev, Z. Fishelson, J. Cuttler, The wall effect in 3he counters, Nuclear Instruments and Methods 71 (3) (1969) 292 – 296. doi:http://dx.doi.org/10.1016/0029-554X(69)90317-6.
  • (11) ISO 8529-1, International Organization for Standardization, Geneva, Switzerland (2001).
  • (12) G. Knoll, Radiation Detection and Measurement, John Wiley & Sons, 2010.
  • (13) R. Batchelor, R. Aves, T. H. R. Skyrme, Helium‐3 filled proportional counter for neutron spectroscopy, Review of Scientific Instruments 26 (11).
  • (14) I. L. Fowler, P. R. Tunnicliffe, Boron trifluoride proportional counters, Review of Scientific Instruments 21 (8) (1950) 734–740. doi:http://dx.doi.org/10.1063/1.1745700.
  • (15) Agostinelli, S. et al, Nucl. Instrum. Meth. A506 (2003) 250–303. doi:10.1016/S0168-9002(03)01368-8.
Comments 0
Request Comment
You are adding the first comment!
How to quickly get a good reply:
  • Give credit where it’s due by listing out the positive aspects of a paper before getting into which changes should be made.
  • Be specific in your critique, and provide supporting evidence with appropriate references to substantiate general statements.
  • Your comment should inspire ideas to flow and help the author improves the paper.

The better we are at sharing our knowledge with each other, the faster we move forward.
""
The feedback must be of minimum 40 characters and the title a minimum of 5 characters
   
Add comment
Cancel
Loading ...
238759
This is a comment super asjknd jkasnjk adsnkj
Upvote
Downvote
""
The feedback must be of minumum 40 characters
The feedback must be of minumum 40 characters
Submit
Cancel

You are asking your first question!
How to quickly get a good answer:
  • Keep your question short and to the point
  • Check for grammar or spelling errors.
  • Phrase it like a question
Test
Test description